13 Technical specifications

Display range

Display

4-digit 7-segment LED display, red, height of digits 14.2 mm
3 LEDs for active sensor
4 LEDs for switching relays

Display range

- 999..9999 (adjustable by user)

Display units (of measurement) with background lighting °C,°F, %

(cut required unit label from sheet and slide it into the slot
on the side)

Input data

Analogue signal input(s)

Measuring range

4 .. 20 mA, sequential (HLB compatible)

Accuracy class

± 0.5 % at 25 °C

Output data

Analogue output

adjustable : 4 .. 20 mA, Ohmic resistance 400 O or

0 .. 10 V, Ohmic resistance 2 kO

Accuracy of the analogue output

± 0.5 % at 25 °C

Rise time

70 ms

Switching outputs

Type

4 relays each with separate common supply

Switching voltage

0.1 .. 250 VAC

Switching current

9 mA .. 2 A

Switching capacity

400 VA, 50 W ((for inductive loads, use varistors)

Life expectancy of the switching contacts

20 million at minimum load
1 million at nominal load

Reaction time (with switching delay = 0 ms)

approx. 20 ms

Setting range of the switching points

1.5 .. 100 % of the pre-set display range

Setting range of the switching hystereses
(switch-back points)

0.5 .. 99 % of the pre-set display range

Switching window upper switch point

2.25 .. 99 % of the pre-set display range

Switching window lower switch point

1.5 .. 98.24 % of the pre-set display range

Interface Serial

interface

RS232

Baud rate 19.200 Bauds – 8 data bits – 1 start and stop bit
– no parity – no handshake

Ambient conditions

Nominal temperature range

0 .. + 50 °C

Operating temperature range

0 .. + 50 °C

Storage temperature range

- 40 .. + 80 °C

- mark

EN 61000 – 6 –1 / 2 / 3 / 4

Other data

Housing

Control panel housing 96 x 48 x 109 mm
Control panel cut-out 92 (+0.8) x 45 (+0.6) mm
Front panel thickness 1.25.. 15 mm
Maximum installation depth 121 mm

Electrical connections

Supply voltage: plug-in terminal block, 2 pole, RM 5.08
Inputs / outputs: plug-in terminal block, 11 pole, RM 3.5
Relays: plug-in terminal block 6 pole, RM 5.08
Max. cross section of the connections 1.5 mm² for inputs /
outputs
Max. cross section of the connections 2.5 mm² for supply
voltage and relays

Supply voltage

85 .. 265 VAC 50 / 60 Hz
or
12 .. 32 VDC

Current consumption: without sensor
HDA 5500-0-2-AC-006

HDA 5500-0-2-DC-006

15 VA for 85 .. 230 VAC

280 mA max. for 24 VDC

Supply of the measurement transmitter

12

±

1 % VDC max. 20 mA per analogue input

Protection class to DIN 40050

IP 20

Reverse polarity protection of the supply voltage,
excess voltage, override protection

provided

Weight approx.

320

g

Advertising