2 overview of basic settings menu – HYDAC HDA 5500-0-2-DC-006 (CM1k) User Manual

Page 8

Advertising
background image

HDA 5500 for HLB 1000 and CS 1000

- 8 -

7.2 Overview of basic settings menu

Menu point

Setting

Setting range

Default setting

Device:

Device selected to display the values measured
by the HLB 1000 connected

Device selected to display the values measured
by the CS 1000 connected

HLB,
CS

HLB

When HLB 1000 is selected

Allocation switching output 1 (S.S.1)

Switching output 1 allocated to the signal

of rel. change in viscosity

Switching output 1 allocated to the signal

of rel. change in dielectric constant

Switching output 1 allocated to the signal

of

rel.

humidity

Switching output 1 allocated to the signal

of

temperature

r.ViS.
r.diE.
r.HuM.
TEMP.

r.ViS.

Switching function switching output 1 (S.M.1)

Switching output 1 set to switch point and

hysteresis mode

Switching output 1 set to switching window mode

SP
Win


SP

Switching direction switching output 1 (S.d.1)

Relay activated

Relay deactivated

on
oFF


on

Switch on delay switching output 1 (T.on1)

Time, in seconds, which must elapse once the particular switch
point has been reached or exceeded in order for a switching
function to occur.


0..99.99 s


0

Switch-off delay switching output 1 (T.oF1)

Time, in seconds, which must elapse once the value has fallen
below the particular switch-back point in order for a switching
function to occur.

0..99.99 s


0

Switching function switching output 2 (S.S.2)

Similar to switching output 1 (see above), and likewise for:

Switching mode (S.M.2)

Switching direction (S.d.2)

Switch-on delay (T.on.2)

Switch-off delay (T.oF.2)

r.ViS.
r.diE.
r.HuM.
TEMP.
SP
Win
on
oFF
0..99.99 s
0..99.99 s

r.ViS.
SP
on
0
0