Configuring analog outputs

You can determine how the voltage of the analog outputs is set.

1 Select "

> Set up > Interfaces > I/O Module > Analog Outputs".

2 For "Config. Analog Output 1" or "Config. Analog Output 2" choose between:

Set by interface

Leakage rate linear

Off

6.3.8.3

Setting the upper scale value for 10 V of the analog output

You can set which leakage rates are represented by 10 V on the analog output. This setting
only comes into effect when "leakage rate linear" is selected for the configuration of the
analog output.

1 Select "

> Set up > Interfaces > I/O Module > Analog Scale".

2 To set the upper scale value, enter a numerical value between, for example, 10 g/a and

1000 g/a using the keys.

6.3.8.4

Setting up the I/O module protocol

Refer to "Interface description HLD6000 (doc. no. kins44e1-a)" for switching between
"ASCII", "LD" (Leak Detection), "Normal" and "Simple" formats.

1 Select "

> Set up > Interfaces > I/O Module > Protocol".

2 Select between "ASCII", "LD", "Normal", and "Simple".

6.3.8.5

Configuring PLC outputs

You can set which function should be displayed on the PLC output.

1 Select "

> Set up > Interfaces > I/O Module > PLC Outputs".

2 Select one of the 8 PLC outputs and allocate a function to it:

Setpoint 1 or 2

CAL internal active

CAL external active

PROOF active

Warning

Error

CAL or PROOF active

CAL request

Run up

Sniffer button

Light barrier

Measuring

Standby

Sniffer is closed

Error or warning

CAL internal possible