2 changing the filter block, 3 sending for maintenance or repairs, Changing the filter block – INFICON HLD6000 Refrigerant Leak Detector User Manual

Page 64: Sending for maintenance or repairs

Advertising
background image

64

8 Maintenance

Tr

an

sl

ation

o

f the

o

rig

in

al

operatin

g

in

str

u

ct

io

ns

HLD

6000

, k

ina4

3

e1-

a

, 140

8

8.2.2

Changing the filter block

Change the filter block with the integrated fine filters at least 1 x per month.

The filter block can be found at the base of the sniffer tip,

see "Fig. 16: The filters in the sniff-

er line", page 63

. To replace the filter block, carry out the following:

1 Switch the device off.

2 Loosen the cap nut at the lower end of the sniffer tip and separate the sniffer tip from

the probe handle.

3 Pull the filter block out of the base of the sniffer tip.

4 Insert a new filter block into position.

5 Attach the sniffer tip back onto the probe handle and tighten the cap nut at the lower

end of the sniffer tip.

Replacement filter block

see "10.1 Accessories and spare parts", page 67

8.3

Sending for maintenance or repairs

You can send your device to INFICON to have it maintained or repaired. For further infor-
mation regarding this topic

see "9.2 Sending in the device", page 65

.