7 Warning and error messages

Tr

an

sl

ation

o

f the

o

rig

in

al

operatin

g

in

str

u

ct

io

ns

HLD

6000

, k

ina4

3

e1-

a

, 140

8

5xx Flow and pressure errors

W543 Flow in the sniffer line is too low

– Filter in the sniffer tip is clogged

– Sniffer tip is clogged or defective

– Replace the filters

– Check the connection of sniffer line to basic unit

(separate and reconnect; try another sniffer line, if
possible).

– If the problem still exists, contact INFICON customer

service.

W544 Valve does not switch back and forth

Internal defect on the sniffer line

– Check the connection of sniffer line to basic unit

(separate and reconnect; try another sniffer line, if
possible).

– If the problem still exists, contact INFICON customer

service.

W545 Flow through measurement line is too

low

– The filter in the sniffer tip is clogged.

– Sniffer tip is clogged or defective.

– Replace the filters

– Check the connection of sniffer line to basic unit

(separate and reconnect; try another sniffer line, if
possible).

– If the problem still exists, contact INFICON customer

service!

W546 Leak on the measurement line

– The filter in the sniffer tip is clogged.

– Leak or defect in the sniffer tip or the sniffer line.

– Replace the filters.

– Check the plug and screw connections.

– Check the connection of sniffer line to basic unit

(separate and reconnect; try another sniffer line, if
possible).

– If the problem still exists, contact INFICON customer

service.

W547 Flow through reference line is too low

– Filter in the sniffer tip is clogged.

– The sniffer tip is clogged or defective.

– Replace the filters.

– Check the connection of sniffer line to basic unit

(separate and reconnect; try another sniffer line, if
possible).

– If the problem still exists, contact INFICON customer

service.

W548 Leak on the reference line

– The filter in the sniffer tip is clogged.

– Leak or defect in the sniffer tip or the sniffer line.

– Replace the filters

– Check the plug and screw connections.

– Check the connection of sniffer line to basic unit

(separate and reconnect; try another sniffer line, if
possible).

– If the problem still exists, contact INFICON customer

service.

W549 Measurement and reference line are

mixed up

– Poor calibration.

– Internal defect on the sniffer line.

– Recalibrate the device.

– Contact INFICON customer service.

Table 10: Warning and error messages (Contin.)

No. Message

Possible

error

sources

Troubleshooting