2 Instrument Description

If the measurements go outside the tolerance band, and the new
measurement, which is outside the tolerance band, is present for at least the
duration of the storage cycle that was set in the active operating mode, then it
is acquired and stored and a new tolerance band is applied.

H

Going outside the tolerance band is not necessarily recorded. If
the measurements return inside the tolerance band within the
storage cycle (t2), then there will be no recording and “Min. storage
cycle” is still active.

10s

10s

10s

10s

10s

10s

10s

10s

t

t1

t2

t3

t4

10s

t6

t5

t1 = storage through “Min. storage cycle” (forced storage) and

application of a new tolerance band.

t2 = no storage, since the measurement is again within the tolerance

at the end of the storage cycle (10sec).

t3 = storage and application of a new tolerance band, since the

measurement has gone outside the tolerance band at the end of
the storage cycle.

t4 = storage and application of a new tolerance band, since the

measurement has gone outside the tolerance band at the end of
the storage cycle.

t5 = storage through “Min. storage cycle” (forced storage) and

application of a new tolerance band.

t6 = no storage, since the measurement is again within the tolerance

at the end of the storage cycle (10 sec).