Function, Adjustment options – Memmert AtmoCONTROL FDA Version 2.X.X User Manual

Page 19

Advertising
background image

D30396 | Date 09/2014

19

AtmoCONTROL FDA Version

Broad parameter representation

parameter representation

Depiction in

icon bar

Mean-

ing

Depiction on

editor thread

Function and

adjustment options

?

15

STAND BY

15

Hold

temper-

ature

Tolerance band

Function

Maintains a set temperature for a

specific time.

Adjustment options

Adjustment options

Name of programme segment

1

Duration (time or infinite ∞)

Temperature to be maintained

Tolerance value above/below

Alarm if limits are exceeded

Safe

2

(For programme example, see page

26)

?

15

STAND BY

15

Change

temper-

ature

0,1 K/min

Tolerance band

Function

Increases or decreases temperature

over a specific time to a set value.

Adjustment options

Adjustment options

Name of programme segment

1

Duration

Target (setpoint) temperature

Tolerance value above/below

SPWT

3

(For programme example, see page

27)

?

15

STAND BY

15

Hold

humid-

ity

5

.

0

%rh

Tolerance band

+

-

Alarm

Safe

on off

70

.0

%rh

12

h

:

00

m

Hold

0

.

0

%rh

Function

Maintains a specific humidity for a

specific time.

Adjustment options

Adjustment options

Name of programme segment

1

Duration (time or infinite ∞)

Humidity value to be main-

tained

Tolerance value above/below

Alarm if limits are exceeded

Safe

2

1

When run, this is displayed in the status bar of the appliance

2

When Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as

specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the

actual values leaves the tolerance band, the clock timer starts again from the beginning.

3

SPWT: Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint

value is reached, even if the set time has already expired. If this is "off", the programme sequence is

continued after the set time has expired, irrespective of whether the setpoint value was reached or not.

Advertising