Function, Adjustment options, Atmocontrol fda version – Memmert AtmoCONTROL FDA Version 2.X.X User Manual

Page 20: Change humid- ity, Name of programme segment, Duration, Target (setpoint) humidity, Tolerance value above/below, Spwt, Hold pres- sure

Advertising
background image

20

D30396 | Date 09/2014

AtmoCONTROL FDA Version

Depiction in

icon bar

Mean-

ing

Depiction on

editor thread

Function and

adjustment options

?

15

STAND BY

15

Change

humid-

ity

2

h

:

30

m

40

.7

%rh

70

.6

%rh

5

.

0

%rh

5

.

0

%rh

+

-

SPWT

on off

Tolerance band

Increase

Function

Increases or decreases humidity

over a specific time to a specific

value.

Adjustment options

Adjustment options

Name of programme segment

1

Duration

Target (setpoint) humidity

Tolerance value above/below

SPWT

3

?

15

STAND BY

15

Hold

pres-

sure

+

-

Tolerance band

Alarm

Safe

on off

hold p

20

mb

10

mb

2

h

:

30

m

500

mb

Function

Maintains a specific pressure for a

specific time.

Adjustment options

Adjustment options

Name of programme segment

1

Duration (time or infinite ∞)

Pressure to be maintained

Tolerance value above/below

Alarm if limits are exceeded

Safe

2

?

15

STAND BY

15

Change

pres-

sure

+

-

Tolerance band

1100

mb

decrease

2

h

:

30

m

500

mb

10

mb

20

mb

SPWT

on off

Function

Increases or decreases pressure over

a specific time to a specific value.

Adjustment options

Adjustment options

Name of programme segment

1

Duration

Target (setpoint) pressure

Tolerance value above/below

SPWT

3

1

When run, this is displayed in the status bar of the appliance

2

When Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as

specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the

actual values leaves the tolerance band, the clock timer starts again from the beginning.

3

SPWT: Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint

value is reached, even if the set time has already expired. If this is "off", the programme sequence is

continued after the set time has expired, irrespective of whether the setpoint value was reached or not.