Memmert ICH/ ICH L Climate chambers User Manual

Page 32

Advertising
background image

32

D30381 | Date 10/2014

Operation and control

Electronic temperature monitoring (

Electronic temperature monitoring (TWW)

TWW)

Electronic temperature monitoring (

Electronic temperature monitoring (

Electronic temperature monitoring (

Electronic temperature monitoring (

The manually set monitoring temperature

min

and

max

of the overtemperature control is

max

max

monitored by an adjustable over/undertemperature controller (TWW) protection class 3.3 acc.

to DIN 12 880. If the manually set monitoring temperature

max

is exceeded, the TWW takes

max

max

over temperature control and begins to regulate the monitoring temperature (Fig. 18).

t

°C

Setting MAX

Set temperature

Controller error

Emergency operation

Fig. 18

Schematic diagram of how TWW temperature monitoring works

Advertising