51

Menu mode

7.5.2 Humidity calibration

Humidity control can be adjusted according to customer requirements by means of three

freely selectable balance points. For each selected calibration point, a positive or negative

compensation correction value between –10 % and +10 % can be set (Fig. 27).

For humidity adjustment, you will need a calibrated reference measuring device.

rh 20

+3,0 %

rh 90

-1,5 %

60%

40%

0%

20%

80%

100%

rh 50

+1,5 %

Fact

ory calibr

ation

Fig. 27

Humidity adjustment (example)

Example: Humidity deviation at 60 % should be corrected.

1. Press the activation key to the right of the

CALIB

display. The display is enlarged

and the temperature adjustment option

is automatically highlighted.

UFP 100

Cal1

°C

-0,2

K

Cal2

100.0

°C

+0,1

K

Cal3

180.0

°C

-0,2

K

JUSTIEREN

Temperatur

PROGRAM

TIME

USER ID

PROTOCOL

PROGRAM

TIME

USER ID

PROTOCOL

letzte Justierung 12.10.2012 12:00

40.0

C

K

0,2

-

C

K

0,1

+

C

K

0,2

-

Calibration

Last updated 12.10.2012 12:00:00

Cal1

Cal2

Cal3

20.0

37.0

5.0

Temperature

Humidity

2. Turn the turn control until

Humidity

is

Humidity

Humidity

highlighted.

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

%rh

%rh

%

0,5

-

%

1,0

+

JUSTIEREN

Temperature

letzte Kalibrierung 12.10.2012 12:00

Cal1

Cal2

10.0

50.0

Humidity

%rh

%

1,0

+

Cal3

80.0

3. Press the confirmation key repeatedly,

until the calibration point Cal2 is selected.

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

%rh

%rh

%

0,5

-

%

1,0

+

JUSTIEREN

Temperature

letzte Kalibrierung 12.10.2012 12:00

Cal1

Cal2

10.0

Humidity

50.0

%rh

%

1,0

+

Cal3

80.0

4. With the turn control, set the calibration

point Cal2 to 60 % rh.

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

%rh

%rh

%

0,5

-

%

1,0

+

JUSTIEREN

Temperature

letzte Kalibrierung 12.10.2012 12:00

Cal1

Cal2

10.0

Humidity

60.0

%rh

%

1,0

+

Cal3

80.0

Advertising