4 timer mode, 5 type of the slide-in unit (grid or shelf), 6 automatic defrosting system (defrost) – Memmert ICH/ ICH L Climate chambers User Manual

Page 46

Advertising
background image

46

D30381 | Date 10/2014

Menu mode

7.3.4 Timer mode

Here, you can choose whether the digital backwards counter with

target time setting (see page 28) should run setpoint-depend-

ent or not. This determines whether the timer should not start

until a tolerance band of ±3 K around the set temperature is

reached (Fig. 25, B) or if it should start right after activation (A).

t

Timer

t

Timer

A

B

°C/°F

t

6 K

Fig. 25

Timer Mode

A

Timer independent of setpoint: Timer starts right after activation

Fig. 25

Fig. 25

B

Timer setpoint-dependent: Timer does not start until tolerance band is reached

A

Timer independent of setpoint: Timer starts right after activation

A

Timer independent of setpoint: Timer starts right after activation

7.3.5 Type of the slide-in unit (Grid or Shelf)

Here, you have to set the type of the slide-in unit (grid or shelf)

used. The selection Shelf enables you to adjust the control

function to the different air flow characteristics in the interior

when using optional sliding shelves instead of the grids that are

part of the standard delivery.

7.3.6 Automatic defrosting system (Defrost)

The integrated automatic defrosting system for the cooling unit

ensures perfect operation of the appliance at low temperatures

and in permanent operation. The defrosting interval can be set in

steps of 6 hours, between 6 and 48 hours. The setting

Off

deactivates automatic defrosting.
Due to long-term operation with a working temperature below

+15 °C or due to a damp chamber load and/or the door opening frequently, ice can form in

the interior over time. Heavy icing may impair the function of the appliance and could dam-

age the cooling system. In this case, the working chamber should be defrosted. This can be

achieved by short heating of the appliance (to 30-40 °C) or by switching it off for a longer

period of time, e.g. over night. The resulting melting water can best be collected with a cloth

at the front edge of the working chamber. Afterwards, the smooth surface of the interior can

easily be cleaned.
Due to this automatic defrosting, there is a minor brief increase in the chamber temperature

at regular intervals. If you would like to further reduce this impairment in performance, you

can shorten the defrosting frequency, e.g. to every 24 hours.
In this case, please observe if there is a permanent drop in the cooling performance or a

strong fluctuation of the actual value, which may be an indication that the cooling unit is icing

over. If this is the case, please set the automatic defrosting system one level higher.

SETUP

Unit

IP address

255.145.136.225

Subnet mask

255.255.0.0

Alarm temp

°C

F

TWW

TWB

Timer mode

Slide-in unit

Grid

Shelf

SETUP

Unit

IP adress

255.145.136.225

Subnet mask

255.255.0.0

°C

°F

Timer mode

Alarm temp

TWW

TWB

Slide-in unit

Grid Shelf

SETUP

Defrost

18h

2/2

Advertising