Malfunctions, warning and error messages

Description

Cause

Action

See

Appliance does not heat

up any more

The mechanical

temperature lim-

iter (TB) perma-

nently switched

off heating.

1. Wait until the appliance

cools down.

2. Reset the TB. To do so,

press the red button on

the right of the appliance

rear side until a clicking

sound can be heard:

If the alarm continues: Contact

customer service

page 2

6.1.2 Humidity monitoring

Description

Cause

Action

See

Error display symbol

HUMIDITY

Set 55

.0%rh

%rh

55

.4

Water tank

empty

Fill the water tank with dem-

ineralised/distilled water and

press the confirmation key

page 21

Alarm display (

MaxAl

)

HUMIDITY

MaxAl

Set 70

.0%rh

%rh

75

.4

Upper humidity

limit exceeded

Open the door for 30 sec. and

wait to see if the appliance

reliably adjusts to the setpoint.
If the error occurs again, con-

tact customer service.

page 2

Alarm display (

MinAl

)

HUMIDITY

MinAl

Set 60

.0%rh

%rh

55

.4

Humidity below

lower limit

Check if the door is closed.
Check the water supply and

the filling level of the water

tank. If required, refill the

water tank.
If the error occurs again, con-

tact customer service.

page 21

page 2