Model 493.07 hpu converter box, About the model 493.07 converter box, Model 493.07 hpu converter box 275 – MTS FlexTest Models 200 User Manual

Page 275

Advertising
background image

Model 493.07 HPU Converter Box

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Digital I/O and Transition Boards

275

Model 493.07 HPU Converter Box

About the Model 493.07 Converter Box

This section describes how to jumper the Model 493.07 Converter Box to
connect a Series 494 Chassis and other controller types to a hydraulic power
unit (HPU). The converter box converts logic-level signals to and from the
Series 494 Chassis to relay signals used by the HPU pump.

For pumps that are 24 V PLC compliant, the converter box is not needed.
This includes all Series 505 HPUs and 506.52–.92 HPUs.

Part numbers

The Model 493.07 Converter Box is available in 24 V DC and 115 V AC
versions for specific controllers.

HPUs that require control voltages that are not compatible with the
controller HPU Interface board require a Model 493.07 Converter Box
designed for the correct voltage.

Connecting a 24 V Converter Box to a 115 V AC HPU can cause damage
and improper operation.

Make sure that the voltage marked on the cover of the converter box matches
the required voltage for the HPU.

* Converter box interfaces to controllers manufactured before 1986

require custom cables.

Model 493.07 Converter Box Part Numbers

C

ONTROLLER

H

ARDWARE

HPU V

OLTAGE

P

ART

N

UMBER

Series 493 and 494

24 V DC

49-969-401

Series 493 and 494

115 V AC

49-969-403

Model 407 Controller

24 V DC

49-969-402

Model 407 Controller

115 V AC

49-969-404

Pre-1986 Controllers*

24 V DC

49-969-405

Pre-1986 Controllers*

115 V AC

49-969-406

CAUTION

Advertising