Model 409 temperature controller connections – MTS FlexTest Models 200 User Manual

Page 186

Advertising
background image

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Model 494.47 Dual UART/Encoder Interface Card

Mezzanine Cards

186

UART Connections for the Model 494.47 Dual UART/Encoder Card

Model 409

Temperature

Controller

JXB

+ Tx/Rx
- Tx/Rx

2
3

I/O Carrier Board

Front-panel

RJ-50 Connectors

JXA and JXB

Model 409

Temperature

Controller

JXA

Mezzanine

Card

494.47

Dual

UART/

Encoder

Conditioner

+ Tx/Rx
- Tx/Rx

2
3

Cable Grounding

1 The cable shield

connects to the metal

shielding on the RJ-

50 cable plug.

2 The cable plug

shielding connects to

the I/O carrier board

body.

3 The I/O carrier board

connects to earth

ground through the

chassis.

Model 409 Temperature Controller Connections

Cable Grounding

1 The cable shield

connects to the metal

shielding on the RJ-

50 cable plug.

2 The cable plug

shielding connects to

the I/O carrier board

body.

3 The I/O carrier board

connects to earth

ground through the

chassis.

I/O Carrier Board

Front-panel

RJ-50 Connectors

JXA and JXB

JXB

2
3
4
7
8
9

+ Tx
- Tx

+ Power
- Power

+ Rx
- Rx

Model 494.05

Handset

JXA

2
3
4
7
8
9

+ Tx
- Tx

+ Power
- Power

+ Rx
- Rx

Mezzanine

Card

494.47

Dual

UART/

Encoder

Conditioner

Model 494.05

Handset

Model 494.05 Handset Connections