MTS FlexTest Models 200 User Manual

Page 136

Advertising
background image

MTS FlexTest® Models 40/60/100/200 Controller Hardware

Mezzanine Cards

Mezzanine Cards

136

494.26

Dual DUC

This card has two digital universal conditioners (DUCs)
that you can independently configure (with software) as
either an AC conditioner or a DC conditioner. Each
conditioner on the DUC card connects to a front-panel
RJ-50 connector on the I/O carrier board.

The AC configuration is typically used to condition an
LVDT. The DC configuration contains circuitry that
supports 1/4-, 1/2-, and full-bridge transducers.

Not required

494.45

8-Input A/D
Card

This card converts up to eight +/- 10 volt analog signals
from external devices to digital signals for use by the
controller.

You can connect the signals directly to the front-panel
RJ-50 connectors or route them through a Model 494.75
8-Input BNC transition board.

Input Signals 1-4 use the top RJ-50 connector (JXA).

Input Signals 5-8 use the bottom RJ-50 connector (JXB).

494.75 8-Input BNC
transition board

494.46

8-Output D/A
Card

This card converts up to eight digital signals to +/- 10 volt
analog signals for use by external devices.

The analog output signals are available from the front-
panel RJ-50 connectors or you can route them through a
Model 494.76 8-Output BNC transition board.

You can also configure your system to route analog
output signals to the Model 494.79 8-Channel Valve
Driver transition board. This board provides drive signals
for up to eight Series 252 Servovalves.

Output Signals 1-4 use the top RJ-50 connector (JXA).

Output Signals 5-8 use the bottom RJ-50 connector
(JXB).

494.76 8-Output
BNC transition
board

494.79 8-Channel
Valve Driver
transition board

Mezzanine Cards (part 2 of 3)