Precrack Parameter Values
Value to Use for Simulation
Parameter
25 mm
Precrack Final Crack Limit
10 Hz
Precrack Frequency
0.100 unitless
Precrack Load Ratio
10%
Precrack Lower Least Squares Fit
Percentage
90%
Precrack Upper Least Squares Fit
Percentage
1000000 count
Precrack Cycle Limit
0.348 kN/mm^1.5
Precrack Final Maximum K
The following table lists the COD (Crack Opening Displacement) parameter values.
COD Parameter Values
Value to Use for Simulation
Parameter
COD
Control Mode
0.03 mm/sec
Ramp Rate
The following table lists the data storage parameter values.
Data Storage Parameter Values
Value to Use for Simulation
Parameter
0%
Precrack Save Percent Limit
100
Store Every Nth Precrack Cycle
20 Hz
CTOD Data Frequency
100 N
Delta Force