0 precrack parameters – MTS Multipurpose Simulated User Manual

Page 45

Advertising
background image

Value to Use for Simulation

Parameter

100 count

Store Every Nth Precrack Cycle

10%

FCG Lower Least Squares Fit
Percentage

90%

FCG Upper Least Squares Fit
Percentage

2000 count

Store Every Nth FCG Cycle

The following list shows test termination parameter values.

Test Termination Parameter Values

Description

Parameter

40 mm

FCG Final Crack Limit

1.000E + 000 mm/cycle

FCG Crack Growth Rate Limit

3 count

Outside Crack Growth Limits
Allowed

10000000 count

FCG Cycle Limit

5.3.2.3.0 Precrack Parameters

The following table lists the precrack parameters and their required values.

Precrack Parameters

Description

Parameter

Specify the desired crack size to achieve in
the precrack activity. When the specified
crack size is achieved, the activity stops.

Precrack Final Crack
Limit

Specify the cycle frequency of the command
signal for the precrack activity.

Precrack Frequency

Specify the ratio of the minimum to maximum
load applied to the specimen. The minimum

Precrack Load Ratio

load is determined by this value and the
maximum load is specified by you.

Specify the percentage of the measured load
range used as the lower limit for linear least

Precrack Lower Least
Squares Fit Percentage

MTS TestSuite 45

5.0 Running Simulation Tests

Advertising