Option 5 – MTS WIFT Mini Transducer Interface User Manual
Page 37
Advertising
TI2SHUNT - Transducer Interface Shunt
SWIFT
®
Mini TI
Software Utilities
37
Option 5
A valid shunt calibration should be performed prior to executing this command.
This option allows an easy means of setting the Shunt Calibration Reference
values after calibration. This is normally done as part of calibration and should
not be done during normal use. After this option has executed, uploads will
contain the new shunt reference values.
Note
This menu choice should only be used by qualified service personnel.