MTS WIFT Mini Transducer Interface User Manual

Page 65

Advertising
background image

Zero the Transducer Interface

SWIFT

®

Mini TI

Installation

65

2. Rotate the adapter plate, as necessary, until the Fz on the axes icon (see the

next figure) printed on the transducer label is pointing up.

3. If not already assembled, attach the inclinometer to the level bracket using

the two 6-32 UNC fasteners provided.

Apply Locktite 222 to the threads on the fasteners. Torque each fastener to 2
N•m (18 lbf•in).

4. Install the inclinometer/level bracket assembly on the transducer by

inserting the dowel pins in the level bracket into the pin pilot holes provided
in the transducer, as shown in the next figure.

The orientation of the inclinometer/level bracket assembly is determined by
the orientation of the anti-rotate bracket.

Axes Icon

S

Setup for Vertical Anti-rotate

Bracket Configuration

Digital Inclinometer

–in this position

should read 0°, ±0.1°

Level Bracket

Mounting Holes (4)

Adapter Plate

Advertising