MTS WIFT Mini Transducer Interface User Manual
Page 65
Zero the Transducer Interface
SWIFT
®
Mini TI
Installation
65
2. Rotate the adapter plate, as necessary, until the Fz on the axes icon (see the
next figure) printed on the transducer label is pointing up.
3. If not already assembled, attach the inclinometer to the level bracket using
the two 6-32 UNC fasteners provided.
Apply Locktite 222 to the threads on the fasteners. Torque each fastener to 2
N•m (18 lbf•in).
4. Install the inclinometer/level bracket assembly on the transducer by
inserting the dowel pins in the level bracket into the pin pilot holes provided
in the transducer, as shown in the next figure.
The orientation of the inclinometer/level bracket assembly is determined by
the orientation of the anti-rotate bracket.
Axes Icon
S
Setup for Vertical Anti-rotate
Bracket Configuration
Digital Inclinometer
–in this position
should read 0°, ±0.1°
Level Bracket
Mounting Holes (4)
Adapter Plate