Operating the TG200

Switching on / off

E

The main switch E is located on the right hand side of the rear panel
of the TG200

Switch on the main switch

ON lights up in the display drying mode

000 lights up in the display temp °C/°F

10 lights up in the display air flow

LED interval off lights up

Two dashes (

- -

) light up in the display time

The values appear at the first start. If the values are
changed they are retained at the end of the programme
and appear when switching on again and pressing the
START key.

The TG200 is now ready for operation, without interval, for
continuous operation without heating and with minimum fan power.

Starting – interruption - stopping

16a

15

16

Starting:

Press START / key 15

Green LED over key 15 lights up

TG200 starts with specified values.

Red LED over key16a flashes as long as a temperature has not

been specified.

Specify temperature, red LED is extinguished, green LED lights

up.

During the drying period the temperature is kept constant within a
tolerance. This means that the heater is switched off when the
specified temperature has been reached and is switched on again
when the temperature drops below the specified value. A
temperature difference of +/- 2°C or +/-50°F is possible here.

Quick start with full power:

Press START / key 15 twice

Green LED over key 15 lights up

TG200 starts with full heating power and fan power for an

indefinite running period

The measured temperature is displayed

Stopping:

Press STOP / key 16 – heater is switched off

Red LED over key 16 flashes

Press STOP / key 16 a second time – everything is switched off

Red LED over key 16 lights up

The complete display except on is extinguished

Activate LED displays = press key 15 once

New values can be entered

Press START / key 15 a second time

Function is carried out as with Starting