Retsch TG 200 User Manual

Page 16

Advertising
background image

2/25/2013

Retsch GmbH

16

0002

Drying mode

Storage and recalling drying parameters

2

5

8

10

11

13

You can store, change or overwright the drying parameters temp –
air flow – interval and time
at any time in the programme
positions P1 to P9.

After switching on the TG200 the indication “on” appears in the
display 2.
By pressing the key PROG the next programme position P1 to P9
can be reached. After programme position P9 on reappears in the
display. In the PROG mode all keys except PROG, START and
STOP
are inhibited.

on = free adjustment of the drying parameters 5/8/10-11
and 13 is possible, or

P1 to P9 = parameters can be saved and recalled here

Start = drying process is started with the parameters
stored in programme position P1 to P9.

Allocation of storage location

Press PROG key until required storage location P1-P9 has
been reached.

Press SET key – all displays flash

Set drying parameters 5/8/10-11/13

Press SET key; values are stored

Flashing of the display stops, the adjustment inhibitor is

activated and sieve parameters are stored. In the PROG

mode all keys except PROG, START and STOP are inhibited.

Signal tone for end of drying process

If the drying process is brought to an end within a definite time
without interruption, the end is signalled by an acoustic signal
tone which is sounded 5 times.

Advertising