RKI Instruments GX-2001 Data Logging with Calibration Station User Manual

Page 46

Advertising
background image

46 • Calibrating the Model GX-2001 with the Calibration Station

8. Connect the polyurethane tubing from the regulator to the

Span Gas In fitting on the back of the GX-2001 Calibration
Station.

9. Make sure the Calibration In and Calibration Out tubing is

properly attached to the Calibration Adapter. This tubing was
connected at the factory.

Auto-Calibrating the Model GX-2001

1. Make sure you have properly setup the GX-2001 Calibration

Station following the instructions under “Preparing for
Calibration.”

2. Turn on the computer, then launch the GX-2001 Data

Logging Software.

3. Check the Automatic Calibration box on the Download

window.

4. Using the On/Off switch, turn on the GX-2001 Calibration

Station.

5. Insert the Model GX-2001 into the Data Logging/Charger

Station on top of the GX-2001 Calibration Station.

Figure 28. Choosing the Automatic Calibration Option

Advertising