RKI Instruments GX-2001 Data Logging with Calibration Station User Manual

Page 49

Advertising
background image

Calibrating the Model GX-2001 with the Calibration Station • 49

Manual Calibration

If the Model GX-2001 is not due for calibration, you can still
calibrate the unit with the Calibration Station using the Cal. Unit
Calibration button on the Download window.

NOTE: If manually calibrating a unit more than once in an hour,

see the Calibration Interval section below for
precautions.

1. Make sure you have properly setup the GX-2001 Calibration

Station following the instructions under “Preparing for
Calibration.”

2. Turn on the computer, then launch the GX-2001 Data

Logging Software.

3. Make sure that the Automatic Calibration box on the

Download window is unchecked, or left blank.

4. Using the On/Off switch, turn on the GX-2001 Calibration

Station.

5. Insert the Model GX-2001 into the Charger/Downloading

Station on top of the GX-2001 Calibration Station.

6. Lock the Calibration Adapter onto the Model GX-2001 into

place by pressing the black button on the right side of the
Calibration Adapter. (Refer to Figure 26.)

7. Turn on the Model GX-2001 by pressing and holding the

Mode button on the Model GX-2001 for about one second.

8. Download information from the Model GX-2001 by clicking

either the Instrument Information button or the Complete
Download button. This action activates the Cal. Unit
Calibration button.