Iv. maintaining your oven – SciGene 400 Hybridization Incubator User Manual

Page 10

Advertising
background image

Model 400 Hybridization Incubator — USER MANUAL

Version 2.1, February 2011 (408) 733-7337, [email protected]

9

IV. MAINTAINING YOUR OVEN

A. Servicing

Turn the power switch to the OFF position and unplug the power cord
before performing any service procedure.

1. Checking and Replacing Fuses

Fuses are located in a removable fuse block below the power cord
receptacle on the back of the unit. Unplug the cord and using a
flat blade screwdriver, remove the fuse block as shown in the
illustration. Remove the fuse holder from the block and gently
remove the two fuses. A blown fuse will appear dark. Always
replace a blown fuse with fuses matching the amperage and
voltage as shown on the label below the fuse block.





















 

  

















 













 





2. Replacing the Temperature Controller

The temperature controller may need to be replaced if the actual
temperature displayed on the controller is erratic after calibration,
or if an error message is displayed. Contact SciGene
(

[email protected]

) for assistance in troubleshooting.


Use small screwdriver behind tab to open Fuse Block

Fuse Block

Push clip up to release Fuse Holder.

Back of Oven

Digital Thermometer Jack

Power

Receptacle

Advertising