V. TROUBLESHOOTING

Symptom

Cause

Solution

Rotator does not turn
after closing door

Rotator control is in the off
position

Turn on rotator.

Door does not shut or is
difficult to open

Latch is improperly
adjusted

Adjust threaded rod
on inside of latch

Temperature is erratic

Thermal controller is
defective

Replace controller

VI. SPECIFICATIONS

Electrical (Oven Unit)
Cat. 1040-60-1 115V AC; 50/60 Hz; 350 W
Cat. 1040-60-2 230V AC; 50/60 Hz; 350 W
Weight

Net

70 lbs (32 kg)

Gross 75 lbs (34 kg) Includes shipping carton

Dimensions (WxDxH)

Interior 14x14x14 inches (36x36x36 cm)

Exterior

18x17x22 inches (45x40x56 cm)

Performance and Controls
Temperature Range

Ambient +5

°C to 99°C

Temperature

Control

± 0.1

°C

Heat

up

Time

3

°C per minute

Temperature Controller

Digital PID, single loop

Temperature Display

Actual/Set dual line LED

Rotation Speed

2 to 20 RPM

Chamber

Stainless Steel

Digital Thermometer Output

Thermocouple

VII. ROTATORS AND ACCESSORIES

Cat. #

Description

UoM

1040-01-0 Hybridization tube, 35x300mm with screw cap

Each

1040-02-0 Hybridization tube, 35x150mm with screw cap

Each

1040-20-5 Replacement Drip Pan for Model 400

Each

1040-21-0 Removable Rocking Platform

Each

1040-21-1 Rotator for 10 large/20 small hybridization tubes

Each

1040-21-2 Maxi-80 Rotator for 80x50ml conical tubes

Each

1040-21-3 Maxi-200 Rotator for 200x15ml tubes

Each

1040-21-4 Maxi-400 Rotator for 400x1.5ml centrifuge tubes

Each

1040-21-5 Maxi-140 Rotator for 40x50mL & 100x15mL plastic tubes.

Each

1040-21-6 Maxi-240 Rotator for 200x1.5mL and 40x50mL tubes.

Each

1040-21-7 Maxi-300 Rotator for 100x15ml tubes & 200x1.5 ml tubes.

Each

1051-52-0 Digital thermometer with NIST certificate of calibration

Each

Advertising