SciGene 400 Hybridization Incubator User Manual

Page 6

Advertising
background image

Model 400 Hybridization Incubator — USER MANUAL

Version 2.1, February 2011 (408) 733-7337, [email protected]

5

III. USING YOUR OVEN

A. Oven Components and Controls

SciGene ovens use a radiant heating system to warm the walls of the
stainless steel interior, providing uniform heating of air within the rotator
compartment. Operating temperature (ambient + 5 to 99°C) is regulated to
±0.2°C from set point by means of a high performance temperature
controller (Watlow EZ-Zone) and a sensitive Resistive Thermal Device (RTD)
used as temperature sensor. The RTD is a stainless steel probe located at
the top of the rotator compartment. SciGene ovens simplify calibration by
incorporating a secondary RTD mounted with the primary RTD.

Key Components:
• Temperature Controller - used to set and observe chamber

temperature

• Power Switch - turns on power to the heater.
• Jog Switch – turns on rotator with door open at selected speed
• Rotator Speed control - turns on and controls rotator speed
• Door Latch –

secures the door closed with button release to open