UVP BioChemi System with EpiChemi II Darkroom User Manual

Page 25

Advertising
background image

C. Chemiluminescent samples (Films and Membranes/Plates)

Emitted light in a light tight environment.

• The time to capture chemiluminescence with the BioChemi system will

depend on several factors, but typically is equivalent to the time it would take
for the same sample on film.

• Always make sure to start integration immediately after the substrate has

been added, especially since some substrates can degrade quickly (we
recommend BioWest Extended Duration substrate).

1.

Move filter wheel to the Clear position.

2.

Open LabWorks software.

3.

Pull down black sample tray and place UVP target or printed text on
the tray.

4. Reset

focus.

5.

Turn overhead light off.

6.

Place sample on the black tray.

7.

Set the zoom by turning the lower portion of the lens all the way to
the right.

8.

Click on the Camera Icon.


9.

The Minimal Dialog window will display. Click Stop/Capt. Button.

Stop/Capt Button





The Options button will become available.

Options Button









WWW.UVP.COM

US Technical Support: 1.800.452.6788

UK Technical Support: +44.0.1223.420022

25