UVP BioChemi System with EpiChemi II Darkroom User Manual

Page 27

Advertising
background image

No Integration – straight accumulation of signal on the chip. The image
captured is representative of all of the information on the chip at the end of
the capture time period.

Application: Acquiring blots with known exposure times.

1.

With the Options box open, click on the Start Preview button.

2.

If not already selected, click on the Acquire tab (On Chip,
Sequential Integration and Dynamic Integration should not be
selected) and adjust the exposure time to the desired amount of
time.

Acquire Button

Capture

3. Set

desired

Exposure Time in minutes, seconds or millisec.

4. Set

Binning: 2 x 2 (horizontal x vertical) recommended

5. Click

Capture.