UVP ChemiDoc-It TS2 Imagers User Manual

Page 5

Advertising
background image

GelDoc-It

TS2

and ChemiDoc-It

TS2

Imagers

5

Specifications

Power Requirements:

100/115V, 50/60Hz; 3.1 Amps at 120 Volts

230V, 50/60Hz; 1.55 Amps at 230 Volts

Mains supply voltage fluctuations are not to exceed 10 percent of the
nominal supply voltage

Pollution Degree:

2

Installation Category:

II

Altitude:

Up to 2000m

Ambient Temperature:

5°C to 40°C

Humidity:

Maximum relative humidity of 80% for temperatures up to 31°C,
decreasing linearly to 50% maximum relative humidity at 40°C


Built-In Touch Screen Computer

Operating System:

Windows

®

7, 64-Bit

Connectivity Ports:

2 USB (side of system)

4 USB, VGA and Ethernet (rear of system)

Wireless Networking Capability:

802.11 b/g/n

USB Flash Drive Capacity:

2 GB (minimum)

Internal Hard Drive:

250 GB (minimum)

Software:

TS2 (Touch Screen) Software (supplied with all
systems)

VisionWorks

®

LS Software (supplied with ChemiDoc-

It

TS2

only; for use on external computer)


Cameras and Lenses

GelDoc-It

TS2

: The GelDoc-It

TS2

is equipped with the GelCam 310 camera, a scientific-grade

monochrome CCD camera with a resolution of 2.0MP (1600x1200). The GelCam 310 is equipped with
a 12.5-75mm f/1.2 optical zoom lens.

ChemiDoc-It

TS2

: The ChemiDoc-It

TS2

is equipped with either the MegaCam 810 or BioChemi™ 510

scientific-grade monochrome CCD camera. Both cameras are Peltier cooled and offer full 16-bit file bit
depth:

The BioChemi 510 camera is a 2.1MP unit with Peltier cooling to 35°C below ambient. The
BioChemi 510 has a peak quantum efficiency of 50% and is capable of binning from 1x1 to 8x8.
This camera is equipped with a 12.5-75mm f/1.2 optical zoom lens.

The MegaCam 810 camera is an 8.1MP camera (3296x2472) with Peltier cooling to 35°C below
ambient. The MegaCam 810 has a peak quantum efficiency of 50% and is also capable of binning
from 1x1 to 8x8. This camera is equipped with either a 50mm f/1.2 lens or a 30mm f/1.4 lens, both
of which have a fixed focal length (non-optical zoom capable).


All camera settings are factory pre-set for optimum performance when viewing gels, films or
membranes under low light level conditions. Contact UVP Technical Support prior to making any
adjustments to the internal camera settings.

Ethidium Bromide (EtBr) Emission Filter

The ethidium bromide (50mm

2

) UV-blocking bandpass interference filter blocks UV and IR radiation

emitted from the transilluminator. The filter is placed in the slide-out filter wheel assembly on the side of
the Imager. The filter allows visualization of fluorophores from 580-630nm, targeting the ethidium
bromide emission peak of 605nm.

Advertising