UVP ChemiDoc-It TS2 Imagers User Manual

Page 6

Advertising
background image

GelDoc-It

TS2

and ChemiDoc-It

TS2

Imagers

6

Fluorescent Focus Target


TS2 Software Screen


Additional filters are available for other specific fluorophores, including custom filters. Filters can also be
removed when imaging non-fluorescent media (including chemiluminescent blots, protein gels, colony
plates, etc.) in order to produce brighter images. Contact UVP for ordering information.

Darkroom

The darkroom is light tight to provide optimal imaging conditions. Darkroom features include:

Epi (overhead) LED white light

Roll-out transilluminator tray

UV-safe gel viewer window built into the darkroom door

Five position emission filter wheel

Brackets and built-in power ports for installing optional epi UV illumination lamps

Brackets for installing optional high-intensity halogen or xenon BioLite™ epi light guides

Built-in power port for transilluminator

Power cord for optional LED White Light Plate

UV safety interlock switch with manual override to disable UV transillumination and
optional UV epi illumination when darkroom door is opened

Transilluminator

The GelDoc-It

TS2

and ChemiDoc-It

TS2

Imagers can accommodate UVP’s Benchtop and FirstLight

®

transilluminator models. UVP offers a variety of transilluminator configurations, including Benchtop
models with multiple wavelengths and variable intensities as well as the highly uniform, patented
FirstLight transilluminator. Refer to the packing slip for the transilluminator included with your system.

Note: For UV protection and to extend the life of the UV transilluminator, the system incorporates a
customizable transilluminator shutoff timer built into the software. For additional information, refer to the
TS2 Software User Manual included with the system or available online at http://uvp.com/manuals.

LCD Touch Screen

The GelDoc-It

TS2

and ChemiDoc-It

TS2

Imagers contain a fully integrated 15.6-inch color touch screen.

The touch screen allows the user to perform a variety of tasks, including previewing, capturing, saving
and printing images, as well as selecting preference options, without the use of an external mouse or
keyboard. The screen can be pressed by finger or with the included pen-shaped stylus.

For users who prefer not to use the touch screen interface, an external keyboard and mouse are
included as standard. These components can be plugged into any available system USB port.

TS2 Software

Image analysis and hardware functions for the GelDoc-It

TS2

and ChemiDoc-It

TS2

Imagers are controlled by the TS2

(Touch Screen) Software interface. In addition to image
preview, capture and save functions, the TS2 Software
controls camera and lens functions (exposure, aperture,
focus and zoom), image printing, lighting and filter selection,
and other user preferences.

For additional information, refer to the TS2 Software User
Manual
included with the system or available online at
http://uvp.com/manuals.

Fluorescent Focus Target

The UVP Fluorescent Focus Target fluoresces when placed
on a UV transilluminator or when exposed to overhead UV.
The Target provides sharp fluorescent images to aid in
adjusting the lens and camera settings for ideal imaging
results.

Advertising
This manual is related to the following products: