Service procedures, Image focusing, Return procedure – UVP GelDoc-It TS Imaging System User Manual

Page 12: Replacement parts and accessories

Advertising
background image

GelDoc-It TS Imaging System

12

Image Focusing

With the Manual Lens

Prior to capturing any images, prepare the image focus capabilities of the system:


1. Turn on the transilluminator and place the Fluorescent Focus Target on the transilluminator surface. NOTE:

The darkroom has a UV safety switch that turns the transilluminator off when the door is open. After closing
the door, be sure the overhead white light switch is turned off.

2. While watching the LCD screen, rotate the lens f-stop adjustment (top ring) so that the image is bright

enough to see on the screen.

3. Rotate the lens focus adjustment (bottom ring) on the lens. Adjust this so that the image appears in clear

focus on the LCD monitor. Note: Once the proper zoom range is set, the lens will have to be refocused
when zoomed in completely.

4. Rotate the zoom lens adjustment (middle ring) on the lens, so that the image is as big as possible. Readjust

the focus ring (bottom ring) on the lens, making the image clear. Adjust the zoom so that the object of
interest is within the picture on the LCD monitor.

With the Motorized Lens

Prior to capturing any images, prepare the image focus capabilities of the system:

1. Turn on the transilluminator and place the Fluorescent Focus Target on the transilluminator surface. NOTE:

The darkroom has a UV safety switch that turns the transilluminator off when the door is open. After closing
the door, be sure the overhead white light switch is turned off.

2. Use the Lighting menu button (under the Acquisition Action Tab) in the VisionWorks software to control the

lens.

Service Procedures

Return Procedure


A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before
returning any product.


Replacement Parts and Accessories


To order accessories or replacement parts for the GelDoc-

It TS System, contact UVP’s offices.

Part Description

Part Number


UV Lamp, UVGL-25, 254/365nm, 4 watt (2 recommended

– 115V)

95-0021-12

UV Lamp, UVGL-25, 254/365nm, 4 watt (2 recommended

– 230V)

95-0021-10

Filter, Ethidium Bromide, 50mm Square

38-0220-01

Filter, SYBR Green, 50mm Square

38-0219-01

Filter, SYBR Gold, 50mm Square

38-0221-01

White Light Tray

95-0476-01

White Light Converter Plate, 21x26cm

38-0191-01

Visi-Blue Converter Plate, 21x26cm

38-0200-01

Gel-Cutter

85-0002-01

Gel-Ruler

85-0003-01

Gel-Scooper

85-0006-01

Gel-Tray, small

85-0007-01

Gel-Sentry DNA Preparation Plate

97-0076-01

Fluorescent Standard Step Tablet

33-0014-02

Spectacles, UV Blocking (UVC-303)

98-0002-01

Goggles, UV Blocking (UVC-503)

98-0002-02

Faceshield, UV Blocking (UVC-803)

98-0002-04