Care and cleaning – UVP HEPA/UV PCR Systems User Manual

Page 11

Advertising
background image

HEPA/UV3 PCR Workstation and Cabinet

Page 11

4. Once the desired time setting is reached, do not press any additional buttons. The timer will

then cycle through the remainder of its settings. When the timer reads “OFF” for 5 seconds
without blinking, the timer is ready to use with the new time setting.


Turn the key to the off position after UV irradiation to prevent exposure of samples to the UV. Ensure
the key is completely inserted into the lock.

[Note] A safety switch sensor is built into the bottom of the door. When the door is opened, the
ultraviolet lights will automatically shut off.

[Note] Do not attempt to perform PCR procedures with the germicidal lamp on. The germicidal lamp is
used to decontaminate the internal chamber of the between experiments.

[Warning] Exposure to UV light is harmful.
The Makrolon

®

door is UV blocking and will not allow UV radiation to pass through the door. The

germicidal tubes will shut off if the panel door is opened. The PCR Workstation contains a powerful
source of UV radiation that will cause damage to unprotected eyes and skin. Before operating any unit,
be sure all personnel in the area are properly protected. Even though the unit shuts the UV off when the
door is open, UV Blocking Eyewear should be worn as well. UVP has a complete line of UV Blocking
Eyewear: Spectacles, Goggles and Faceshield designed for this purpose.

Extended periods of time UV exposure can be used to disinfect the interior chamber by the destruction
of organisms. Refer to the Maintenance section for information on measuring the intensity of the UV
tubes with a UV meter.


Overhead White Light

The overhead white lights can remain lit at any time throughout experiment procedures or the
decontamination process. The fluorescent white light provides a bright light within the PCR Workstation.
To use the white light, press the ON button located on the top panel.

Care and Cleaning

The units are built to provide trouble-free operation.

NOTE: The door and side panels are made of Makrolon

®

and are subject to scuffing and scratches if

improperly cleaned. If crazing of the Makrolon

®

panel occurs, the panel can be replaced. Refer to

Replacement Parts for part numbers.

The stainless steel and aluminum surfaces are manufactured with an antimicrobial coating to reduce
bacterial growth. Care in cleaning and use of the equipment is recommended to reduce wear of the
coating. To clean the unit:

Wipe excess water from inside the unit and outside the unit with an absorbent soft cloth or
sponge.

Use mild soap and a damp soft cloth or damp sponge to clean the exterior of the unit.

Use a mild detergent on the interior surfaces.

Warning: Unplug the unit before cleaning around the UV tube contacts and internal plug
outlets.

Clean the door with a mild detergent; never use organic based compounds, Alcohol, or
Ammonia containing cleaners.

Do not use abrasive pads or cleansers.

A plastic cleaner solution is recommended to clean the door and is available from local plastic
supply distributors.