UVP HEPA/UV PCR Systems User Manual

Page 13

Advertising
background image

HEPA/UV3 PCR Workstation and Cabinet

Page 13

1. To replace any of the filters or the UV tube in the

HEPA/UV system, unscrew knobs (2) located on
the ceiling of the workstation to release
compression in the HEPA/UV system.


2. Open the HEPA/UV door located on the left of the

workstation to access the filters.

3. Pre-filter. The Pre-filter (38-0314-01) is located on

the inside of the HEPA/UV door. To remove, open
the door and slide the filter off of the door.

4. Carbon Filter. Slide the filter assembly out of the chamber. The Carbon Filter (76-0352-01) then

slides to the right out of the filter assembly. If the filter does not slide out easily, loosen the thumb
screws in front of the filter.

5. HEPA Filter. To replace the HEPA filter, slide the filter assembly out of the chamber and

remove the carbon filter if not already removed. Turn the knobs counterclockwise to loosen.
Open the HEPA door.


Slide the HEPA filter out and replace filter. Close door and tighten the knobs (turn clockwise). Insert
the carbon filter. Tighten the carbon filter thumbscrews if loose. Close filter door. Retighten the
knobs (2) on the workstation ceiling.


Knobs

UV Tube

Pre-Filter

Carbon

Filter

HEPA

Filter

Thumb
screws