HEPA/UV3 PCR Workstation and Cabinet

Page 3

HEPA/UV

Three stage filter
UV Source

Pre-filter, carbon-filter and HEPA-filter
One 8-watt 254nm shortwave UV tube

Chamber Light Sources

Workstation:
UV Source - Chamber
UV Air Circulator
White Light

Cabinet:
UV Source - Chamber
UV Air Circulator
White Light


Two 25-watt 254nm shortwave UV tubes
One 8-watt 254nm shortwave UV tube
Two 8-watt white tubes


Four 8-watt 254nm shortwave UV tubes
One 8-watt 254nm shortwave UV tube
Two 8-watt white tubes

Power Outlets

Built-in outlets for operating instruments inside the chamber:
Four in the Workstation, two in the Cabinet

Shelves

Construction
Dim. (for Workstation)
Dim. (for Cabinet)

Two - Formed antimicrobial coated aluminum
13”W x 4.2”D (330 x 107mm)
8.75” x 4”D (222 x 102mm)

Construction

Interior

Antimicrobial coated stainless steel (horizontal work surface)
and aluminum

Exterior

Aluminum powder coated

Door and Side Panels

Two panel Makrolon

®

assembly

Blocks UV under 400nm

Electrical

115V - 60 Hz, 230V - 50 Hz, 100V - 50/60Hz

Operational Ratings

Unit intended to be used indoors

Altitude must not exceed 2000 m

Ambient temperature not to exceed 5°C to 40°C (41°F to 104°F)

Relative humidity not to exceed 80% for temperatures up to 31°C (88°F) decreasing linearly to 50%
relative humidity at 40°C (104°F)

Mains supply voltage fluctuations up to ±10% of the nominal voltage

Pollution rating 2 or better for laboratory areas

Part
Number

Voltage

Max
Amps

95-0434-01
95-0434-02
95-0434-04
95-0434-03
95-0434-05
95-0438-01
95-0438-02
95-0438-04
95-0438-03
95-0438-05

115V
230V
230V
100V
230V
115V
230V
230V
100V
230V

9.1
8.7
8.7
9.1
8.7
9.1
8.7
8.7
9.1
8.7