Specifications

SECTION 7

Test Sample: Whole blood

Test Result: Plasma/Serum glucose

Reaction Chamber

Sample Volume: 2

µ

L

Measuring Range: 20–600 mg/dL (1.1–33.3 mmol/L)

Measuring Time: 30 Seconds

Calibration Curve: Selected by use of the Code Strip

Memory Feature: Stores last 20 test results

Temperature Compensation: Automatic compensation with

built-in thermistor

Battery Type: 3-volt lithium batteries (DL or CR2032) – 2

Battery Life: Approximately 1000 Tests (1 year of average use)

Operating Temperature Range: 50°–104° F (10°– 40° C)

Humidity: 20–80% RH

Dimensions: 81 x 51 x 14 mm (3

11

32

x 2 x

9

16

in.)

Weight: Approximately 50g (1.75 ounces)

Alarm: “Beeps” whenever a Check Strip, Code Strip or Test

Strip is inserted into the Meter, when the Test Strip
is completely filled with blood or Control Solution,
and when a test result appears in the Display.