Calibration

9

Chapter 4

Calibration

Every calibration procedure will finish by providing a table of voltages
that correspond to power levels at the calibration frequency (f

0

). It will

be up to the user to use those voltages to generate an equation or fill in
a look up table on whatever they choose to use as an output.

Zero Calibrating

Note:

This procedure can be performed any time the RF is

off and the TPM is on. There is no need to cycle power again on
the TPM.

Do one of the following:

Locate pin 9 on the TPM’s DB9 connector and short it to ground.

If connected to a display panel, press the zero calibration
switch. Refer to Figure 4 for Zero Calibration Switch location.

Note:

Zero calibrate will set both the FWD and RFL at the

same time.

CAUTION

A Bird Calibration Kit is recommended for this procedure.

Performing a calibration with anything other than a Bird

Calibration Kit voids Bird’s accuracy guarantee.

Performing a field calibration will void any remaining warranty

on this product. Bird recommends waiting until the terms of the

warranty have passed before performing a field calibration.

1

6

9

5

Pin 9

Figure 4

Pin 9