tenuation require allowance for normal instrument er-

rors. The correction may be determined by direct rigid

connection of the Thruline Wattmeter units, or of the

Thruline Wattmeter - Termaline Load combination, in cas-

cade. Slight juggling of the zero settings is permissible

for convenience in eliminating computation, provided read-

ings are being taken fairly well up on the scale.

3. Attenuation By Open Or Short Circuit Method. More

clever by far than method (2) is one depending on the

high directivity (null balance) to which the Thruline Watt-

meter elements are held. They should, and do, exhibit

good equality between forward and reflected readings

when the load connector is open or short circuited. In

this condition,

φ

= 100 percent, the forward and reflected

waves being equal in magnitude, and p = infinity. Say

that this is checked on open circuit, and then a length of

line of unknown attenuation, also open circuited, is con-

nected to the load connector. The ratio

φ

then shown is

the attenuation in two passes along the line (down and

back).

Expressed in dB,

the dB figure may be compared with published data for line type and length

by remembering to halve N

dB

or to double the line length, because twice

the line length is actually being measured.
This measurement should be supplemented by one of

φ

(Insertion), or at least

by dc continuity and leakage checks, since the attenuation measurement

alone can be in error from faults such as open or short circuits part of the

way down the line.

using the equation: N

,

dB

=

F

HG

I

KJ

10 log

W

W

r

f

Theory of Operation

15