1. Insert the appropriate element in the socket of the line sec-

tion.

2. Rotate the element so that the “arrow” on the nameplate

is pointed away from the RF source for forward power

and towards the source for reflected power.

3. Turn on the RF source.
4. Read the power level indicated on the appropriate meter

scale.

In cases where readings are being made when the meter unit is connected

to an auxiliary RF line section, always remove any measuring element from

the unused RF line section. Otherwise, the dc circuit will be unbalanced or

shorted according to the “arrow” position of the other element, causing

inaccurate or no indication on the meter.
For the convenience of users, a set of VSWR conversion nomographs is

included in this instruction book. With these charts, VSWR may be directly

determined from forward and reflected power values read from the Thruline

Wattmeter.
When a Model 43 is used to match a load to a transmitter and a good match

is obtained, removing the instrument will not cause any change in the