will also be noticed upon transmitter de-energization as the meter needle

sweeps back to zero. Also, the meter reading may be expected to waver

approximately 2-3 minor scale divisions when voice modulation is used as

the peak detector attempts to follow peak power variations associated with

voice modulation.

Model 43P Peak Calibration

If it is felt that the peak readings are questionable, or for any reason

recalibration is required, proceed as described.
Remove the rear cover, follow these instructions.

1. Remove the four 8-32 Phillips flat head screws that se-

cure the rear cover. There are two on either side of the

meter housing.

2. Pull the cover straight off.

Refer to figure 15 for the following calibration instructions:

1. Insert the Model 43P between an RF (CW) signal source

and a 50 ohm load.

2. Insert an element for the appropriate power and fre-

quency range into the Model 43.

3. Turn the element in the direction to read forward power.
4. Place the mode switch, on the side of the Model 43P, in

the CW position (LED off).

5. Turn on the signal source.
6. Adjust power to give a stable reading on the Model 43P

Wattmeter in the upper half of the scale.

7. Switch to the PEAK mode (LED on).
8. The RF power reading should remain the same. If it does

not read the same, adjust potentiometer (R9 -top center

of the PC board) until the PEAK reading is the same as

Bird Model 43 Wattmeter