PC-2200

BOWERS METROLOGY page 12 /32 version 2.06 -E



3.2 PROGRAMMING A SIMPLE MEASURE

Example N°1
Measure a component with a HBT probe connected to channel A.


Step n°1 is to place the probe into a correct position to be able to get the maximum range and the minimum
deviation.
Place the master 20.0000 mm on the stand, place the probe into position.
Press simultaneously

Move the probe down to the electro-mechanical zero, , then approximately stop around

zero, lock the probe . Press to escape to the measuring mode.





Parameters to be changed for example number 1

MEASURE : M1 –Range :

± 0.01mm Resol.:0.0002mm

SPEC.

: M1 – NOMI. 020.0000 - -TOLER –0.0050 - +TOLER +0.0050 –APPROCH –0.0040

+

APPROCH

:

+0.0040



Master

20.0000mm

Component

20.000mm ± 0.005

USING a HBT probe, we do not use the
calibration, this mode will be explained in a
subsequent chapter.

S

Enter

Enter

Esc

A D J U S T

-

0 . 0 0 1 0

C H - A

- 1 . 0 5 0 0