PC-2200

BOWERS METROLOGY page 27 /32 version 2.06 -E


5.2 HOW TO SETUP FOR A/E CONVERTER




A.

BALANCE

B

.

PRESSURE




E.

SIGNAL



C.

GAGE

D.

AIR IN





It is necessary to adjust the A/E CONVERTER in order to use this Column optimally.


1) Pressure Adjustment(STANDARD : 2 ± 0.1bar)

-

Connect PRESSURE GAUGE to “C”.

-

Connect AIR(Min. 3 bar) to “D”.

-

Adjust pressure by turning “B” with screw driver.


♦ Default pressure is 2 bar(Factory setting).




2) Balance

adjustment

-

Connect the A/E converter and the Column with cable(PC-2332).

-

Connect the air gauge to “C”.

-

Stop supplying “D” with air.

-

Preset the Column to “0” with the “PRESET” key.

-

Place the air gauge into the Min. Master.

-

Supply “D” with air.

-

Adjust “A” with screw driver.

) Example) Min. Master value : -0.020mm

DISPLAY VALUE : -0.020mm


In case of unbalance, measuring range could be narrow.