Metrohm 726 TP Backup User Manual

Page 25

Advertising
background image

726 TP Backup

24

8.726.8007 Instructions for Use

During the transfer (receive)

This message appears if no handshake is used and the transmit buffer is too
small, for example because the PC cannot process (save) the incoming data at
the present time. Confirm the message with OK, the transfer will be terminated.
The transfer must also be aborted on the 726 Titroprocessor. To do this, switch
to the dialog page Config. / Interfaces and press – possibly several times – the
softkey [Reset RS232 1] or [Reset RS232 2].

It is preferable to work with a handshake. If this is not possible for some reason
or other, you should increase the size of the transmit buffer and ensure that the
PC is free to receive the data during the transfer.

This message can have two causes: either the transfer was started at the
726 Titroprocessor before the PC was set to receive or inadmissible reports were

sent.
Terminate the transfer (Stop) and reset the RS232 interface of the
726 Titroprocessor. To do this, switch to the dialog page Config. / Interfaces
and press – possibly several times – the softkey [Reset RS232 1] or [Reset
RS232 2]
.

In a transfer from the 726 Titroprocessor to the PC, ensure that you do not
send data from the 726 Titroprocessor before you have clicked Receive.