6 Troubleshooting

940 Professional IC Vario ONE/ChS/LPG (2.940.1250)

■■■■■■■■

95

Problem

Cause

Remedy

Replace the separation column (see "Con-
necting the separation column", page 53)
.

Precision problems -
the measured values
are highly scattered.

Injection valve – Sample
loop.

Inspect the installation of the sample loop (see
"Optional: Exchanging the sample loop",
page 28)
.

Sample – There are gas
bubbles in the sample.

Use the sample degasser .

Sample – The rinsing vol-
ume is too small.

Increase the rinsing time (see Chapter 5.14,
page 88)
.

Injection valve – Defective.

Request Metrohm Service.

Vacuum is not being
built

Eluent degasser – Vacuum
connection on the rear of
the instrument is not
(tightly) sealed.