Metrohm 940 Professional IC Vario ONE/ChS/LPG User Manual
Page 103
Advertising
■■■■■■■■■■■■■■■■■■■■■■
6 Troubleshooting
940 Professional IC Vario ONE/ChS/LPG (2.940.1250)
■■■■■■■■
95
Problem
Cause
Remedy
■
Replace the separation column (see "Con-
necting the separation column", page 53).
Precision problems -
the measured values
are highly scattered.
Injection valve – Sample
loop.
Inspect the installation of the sample loop (see
"Optional: Exchanging the sample loop",
page 28).
Sample – There are gas
bubbles in the sample.
Use the sample degasser .
Sample – The rinsing vol-
ume is too small.
Increase the rinsing time (see Chapter 5.14,
page 88).
Injection valve – Defective.
Request Metrohm Service.
Vacuum is not being
built
Eluent degasser – Vacuum
connection on the rear of
the instrument is not
(tightly) sealed.
Advertising