17 installing the low-pressure gradient – Metrohm 940 Professional IC Vario ONE/ChS/LPG User Manual

Page 51

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

940 Professional IC Vario ONE/ChS/LPG (2.940.1250)

■■■■■■■■

43

3.17

Installing the low-pressure gradient

Accessories

For this step you need the following accessories:

Mixing coil for low-pressure gradient (6.2758.020)

2 × aspiration tubing, 2 m (6.1834.080)

Installing the low-pressure gradient

1 Removing the eluent degasser – high-pressure pump connec-

tion tubing

Loosen the clamping screw at the output of the eluent degasser
and remove the connection tubing.

Loosen the pressure screw on the high-pressure pump's inlet
valve holder and remove the connection tubing, the coupling and
the pump head input capillary.

2 Removing the stoppers

Remove the stopper at eluent input A.

Remove the stopper from the eluent output (labeled Out).