9 injection valve, 10 suppressor, 11 detector – Metrohm 940 Professional IC Vario ONE/ChS/LPG User Manual

Page 107: 12 sample degasser, Injection valve

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Technical specifications

940 Professional IC Vario ONE/ChS/LPG (2.940.1250)

■■■■■■■■

99

Gradient capabil-
ity

Isocratic or gradient (extendable to quaternary)

Profile

Step, linear, convex and concave

Resolution

< 1 nL/min

7.9

Injection valve

Actuator switch-
ing time

typ. 100 ms

Maximum operat-
ing pressure

35 MPa (350 bar)

Material

PEEK

7.10

Suppressor

Resistance to sol-
vents

No restriction

Switching time

typ. 100 ms

7.11

Detector

You can find the technical specifications for the detector in the detector's
manual.

7.12

Sample degasser

Material

Fluoropolymer

Resistance to sol-
vents

No restriction (except PFC)

Time to establish
vacuum

< 60 s

Advertising