Metrohm 850 Professional IC AnCat MCS User Manual

Page 111

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Troubleshooting

850 Professional IC – AnCat – MCS

■■■■■■■■

101

Problem

Cause

Remedy

Eluent – Evaporation of the
organic solvent in the elu-
ent.

Check the eluent bottle attachment (see Figure
13, page 29)
.

Peak areas lower
than expected

Sample – Leak in the sam-
ple path.

Check the sample path.

Sample – Blockage in the
sample path.

Check the sample path.

Sample – Sample loop not
(completely) filled.

Prolong the sample transfer time.

Sample – Gas bubbles in
the sample.

Use the sample degasser (see Chapter 2.13,
page 40)
(if available).

MCS – not connected.

Connect the MCS.

Peristaltic pump –
insufficient or no
delivery rate

Peristaltic pump – contact
pressure too weak.

Correctly set contact pressure (see "Set flow
rate", page 54)
.

Peristaltic pump – filter
blocked.

Replace the filter (see Chapter 4.12.2.2, page
95)
.

Peristaltic pump – pump
tubing defective.

Replace pump tubing (see Chapter 4.12.2.1,
page 94)
.

Very noisy baseline

High pressure pump – con-
taminated pump valves.

Clean pump valves (see Chapter 4.5.2, page
73)
.

Eluent – Leak in the eluent
path.

Check the eluent path.

Eluent – Blockage in the
eluent path.

Check the eluent path.

High pressure pump –
defective piston seals.

Replace piston seals (see Chapter 4.5.2, page
73)
.

MCS –
CO

2

adsorption cartridge

exhausted.

Replace CO

2

adsorption cartridge (see Chapter

4.13.1, page 96).

Pulsation damper not con-
nected.

Connect the pulsation damper (see Chapter
2.12, page 38)
.

Pulsation damper not con-
nected.

Connect pulsation damper (see Chapter 2.12,
page 38)
.

Advertising