Metrohm 850 Professional IC AnCat MCS User Manual

Page 112

Advertising
background image

5.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

102

■■■■■■■■

850 Professional IC – AnCat – MCS

Problem

Cause

Remedy

MCS – vacuum pump
defective.

Contact the Metrohm Service.

Background conduc-
tivity too high

MSM – not connected.

Connect the MSM (see Chapter 2.16, page
47)
.

MCS – not connected.

Connect the MCS.

Incorrect eluent.

Change eluent (see Chapter 4.4.2.3, page 72).

MSM – regeneration or
rinsing solution flow prob-
lems.

Check the flow of the regeneration and rinsing
solution (see Chapter 2.16.2, page 47).

Data of the separa-
tion column cannot
be read.

Column chip contamina-
ted.

Clean contact surfaces of the column chip
(with alcohol).

Column chip defective.

1. Save column configuration in MagIC Net.
2. Notify Metrohm Service.

Individual peaks
greater than expec-
ted

Sample – Carry-over of the
samples from previous
measurement.

Rinse system longer between two samples.

MSM – No (or insuf-
ficient) pumping of
regeneration or rins-
ing solution

Leak in the system.

Check connections.

Peristaltic pump – contact
pressure too weak.

Correctly set contact pressure (see "Set flow
rate", page 54)
.

Peristaltic pump – filter
blocked (see Figure 29,
page 52).

Replace the filter (see "Replacing the filter",
page 95)
.

MSM – backpressure too
high.

Clean the MSM (see Chapter 4.11.3.2, page
89)
or replace parts (see Chapter 4.11.3.3,
page 92)
.

Peristaltic pump – pump
tubing defective.

Replace pump tubing (see Figure 28, page 51).

Poor reproducibility
of the retention
times

Eluent – Leak in the eluent
path.

Check the eluent path.

Eluent – Blockage in the
eluent path.

Check the eluent path.

Chromatograms
have poor resolution

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 4.15.4, page 98)
.

Advertising