1 wavelength linearization, Wavelength linearization – Metrohm NIRS XDS Transmission OptiProbe Analyzer User Manual

Page 40

Advertising
background image

38

▪▪▪▪▪▪▪

This is used only on certain XDS Process Instruments. Do not select for XDS Smart Probe.

When calibration or library transfer (between instruments) is anticipated, be sure that
Instrument Calibration is selected. This feature assures best method transfer between
similar instruments.

Some of these selections are spectroscopic
calibrations, and are used to apply corrections to
the instrument to minimize differences between
units of the same configuration.

The correction programs are accessed from the
Diagnostics menu bar, shown at right. The
diagnostic steps are explained in the sections that
follow.

When Vision first communicates with the
instrument (or after disconnection and re-
connection) this screen is shown.

Vision must establish the configuration and
have it verified by the user. This assures
that test data is sent to the correct location
in the Diagnostic Database.

If the Sample Module Serial Number field is
empty, locate the sample module serial
number on the side of the module. Record
the serial number and enter it in this field.

Click “OK to accept the instrument
identification.

7.1.1

Wavelength Linearization

Wavelength Linearization uses internally-mounted wavelength materials to determine a set of
internal, arbitrary peak positions that the instrument will use to maintain repeatability of wavelength
response.