3.4 Mounting the sample insert

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

860 KF Thermoprep

3.4

Mounting the sample insert

6.2063.010

6.2627.000

Figure 9

Mounting the sample insert

Note

The sample insert may not be inserted or removed unless it has been
cooled down.

The dimensions of the supplied sample insert are optimized for the usage
of 6 mL sample vials 6.2419.007 from Metrohm. This ensures the best
possible transfer of heat between oven and sample.

1 Place the 6.2627.000 oven insert into the oven from above.

2 Place the 6.2063.010 sample insert into the oven from above.

If vessels with other dimensions are to be used, then individually modified
sample inserts can be ordered. The precise vessel dimensions (including
tolerances) will be required when ordering. Non-standard sample vessels
may not exhibit dimensions outside of the following limit values:

Diameter 10.0…32.0 mm

Immersion depth 20.0…45.0 mm