4 Operation
860 KF Thermoprep
■■■■■■■■
27
Only when 'gas type' = 'other'.
Range
0.001 ... 2.000 (Increment: 0.001)
Default value
1.000
Table 3
Measuring factors of different gases
Gas type
Measuring
factor
Argon
0.950
Helium
0.300
Oxygen
1.000
Air
1.000
Nitrogen
1.000
Flow source
Selecting the flow source.
Selection
pump | valve
Default value
pump
pump
Built-in air pump.
valve
Inlet valve for inert gas.
4.6
Configuration
[TEMP]
The basic settings for the instrument can be made in the Setup menu. You
can call it up during switch on by holding down the [TEMP] key. Press
the [QUIT] key to exit the menu.
Note
The Setup menu can also be called up when the keylock is switched on.
Lock keys
Keylock. The [OK],[
◀] and [▶] keys can be locked. You can thus prevent
changes from being made in the Parameters menu or of the gas flow rate
during a determination. If the function is switched on (on), then the only
switching that can be done is between Temperature 1 and Temperature 2,